Paper · 1985-03 · IEEE Transactions on Biomedical Engineering

A Real-Time QRS Detection Algorithm

Jiapu Pan, Willis J. Tompkins

Work 2 of 2 dated on the heartbeat-counting line, 9 years after the first documented work (US4295472A).

Abstract

We have developed a real-time algorithm for detection of the QRS complexes of ECG signals. It reliably recognizes QRS complexes based upon digital analyses of slope, amplitude, and width. A special digital bandpass filter reduces false detections caused by the various types of interference present in ECG signals. This filtering permits use of low thresholds, thereby increasing detection sensitivity. The algorithm automatically adjusts thresholds and parameters periodically to adapt to such ECG changes as QRS morphology and heart rate. For the standard 24 h MIT/BIH arrhythmia database, this algorithm correctly detects 99.3 percent of the QRS complexes.

Read the paper